Clinical evaluation of Synthetic Aperture Sequential Beamforming and Tissue Harmonic Imaging
Publikation: Bidrag til bog/antologi/rapport › Konferencebidrag i proceedings › Forskning › fagfællebedømt
This study determines if the data reduction achieved by the combination Synthetic Aperture Sequential Beamforming (SASB) and Tissue Harmonic Imaging (THI) affects image quality. SASB-THI was evaluated against the combination of Dynamic Received Focusing and Tissue Harmonic Imaging (DRF-THI). A BK medical UltraView 800 ultrasound scanner equipped with a research interface and an abdominal 3.5 MHz 3.5CL192-3ML convex array transducer was used and connected to a stand alone PC. SASB-THI and DRF-THI scan sequences were recorded interleaved and processed offline. Nineteen patients diagnosed with focal liver pathology were scanned to set a clinical condition, where ultrasonography is often performed. A total of 114 sequences were recorded and evaluated by five radiologists. The evaluators were blinded to the imaging technique, and each sequence was shown twice with different left-right positioning, resulting in 1140 evaluations. The program Image Quality Assessment Program (IQap) and a Visual Analog Scale (VAS) were applied for the evaluation. The scale ranged from -50 to 50, where positive values favored SASB-THI. SASB-THI and DRF-THI were evaluated alike in 49% of the evaluations, 28% favored SASB-THI and 23% favored DRF-THI. The average rating was 0.70 (Cl: -0.80 to 2.19). The statistical analysis, where the hypothesis of no differences between the techniques was tested, yielded a p-value of p=0.64, indicating no preference to any technique. This study demonstrates that SASB-THI and DRF-THI have equally good image quality although a data reduction of 64 times is achieved with SASB-THI.
Originalsprog | Engelsk |
---|---|
Titel | IEEE International Ultrasonics Symposium, IUS |
Antal sider | 4 |
Forlag | IEEE Computer Society Press |
Publikationsdato | 20 okt. 2014 |
Sider | 1312-1315 |
Artikelnummer | 6931774 |
ISBN (Elektronisk) | 9781479970490 |
DOI | |
Status | Udgivet - 20 okt. 2014 |
Begivenhed | 2014 IEEE International Ultrasonics Symposium, IUS 2014 - Chicago, USA Varighed: 3 sep. 2014 → 6 sep. 2014 |
Konference
Konference | 2014 IEEE International Ultrasonics Symposium, IUS 2014 |
---|---|
Land | USA |
By | Chicago |
Periode | 03/09/2014 → 06/09/2014 |
Navn | IEEE International Ultrasonics Symposium, IUS |
---|---|
ISSN | 1948-5719 |
ID: 239625772